The effective area of a piston gauge is measured by either the direct balance or the method of initial balance. Herein, by analyzing and comparing the measurement of direct method and the other method which makes material pressure distortion coefficient and temperature compensated, it is found that the results of the two methods are different based on the different types of piston gauges.