Abstract:During repeated shock accelerometer calibration, it’s found that different mounting of sensor on the comparison-method calibration device will result in different calibration results, and may even cause severe deterioration on the linearity of the measured sensitivity magnitude. This paper systematically analyzes the reasons how the different sensor’s installation result in different testing results, and proposes the corresponding solutions. The analysis here is important to standardize and improve the calibration process.