Application of Standard Line-scale to Calibration of Indentation Diameter Measurement System
DOI:
CSTR:
Author:
Affiliation:

Clc Number:

TH711

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    During the high accurate measurement,it needs to calibrate the value of the CCD pixel in order to fit the accuracy required.In this paper a simple and convenient method for calibrating the CCD pixel with standard line-scale is introduced.The line-scale is calibrated by laser interferometer.We can get the number of pixel corresponding to a length of standard line-scale.Then we can get the real distance corresponding to every pixel.The results of the experiments show that this method can restrain the composit...