Calibration of Plural Sensitivity of Hydrophones Using Laser Method
DOI:
CSTR:
Author:
Affiliation:

Clc Number:

TB565.1

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    The laser vibration measuring technique can be used to directly measure the water particle vibration velocity caused by the sound field,then the amplitude and phase of acoustic pressure are derived from the measurement.Thereby,the plural sensitivity of the hydrophones can be accurately calibrated.Using this technique,we establish a calibration facility for plural sensitivity of hydrophones within the frequency range from 10kHz to 100kHz.This article introduces the calibration principle,method and every part...