Measurement Uncertainty of Pulse Width of Pulse Generator
DOI:
CSTR:
Author:
Affiliation:

Clc Number:

TM930.115 TM935.41

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    In this paper,the uncertainty of evaluation results of pulse width of pulse generator is described.The main error sources of pulse width measurement,including counter's repeatability,trigger level timing error,time base error,the amplifier hysteresis error,LSD displayed error,etc.are analyzed.According to these effect factors,the combined standard uncertainty and expanded uncertainty of pulse width measurement are evaluated.