毫米波频段液体及粉末的介电常数测试技术研究
CSTR:
作者:
作者单位:

作者简介:

通讯作者:

中图分类号:

基金项目:


Millimeter-wave dielectric constant measurement techniques for liquids and powders
Author:
Affiliation:

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    为准确测试毫米波频段液体及粉末材料的介电常数,提出基于波纹传输线的测试方法,利用聚四氟乙烯材质的特制容器搭建测试系统,结合矢量网络分析技术开展测试。首先通过时域门控-反射-传输线(Gated-Reflect-Line, GRL)方法进行校准,降低容器与待测材料间多次信号反射、散射产生的误差影响,之后放入待测液体或粉末材料,采用矢量网络分析仪获取反射参数S,进而计算得出介电常数。以固体聚四氟乙烯为标准样品进行测试,验证了该系统的应用可行性;对石英粉末及二甲基合成硅油进行测试,在75 ~ 110 GHz频段内,测试结果偏离平均值的最大误差仅为5%,说明该测试方法具有较高准确性,为食品加工、生物医学等领域中毫米波频段液体及粉末材料的介电常数测量提供了重要技术借鉴。

    Abstract:

    To achieve accurate dielectric constant measurement of liquid and powder materials in the millimeter-wave band, this paper proposes a measurement method based on corrugated transmission lines. A dedicated test system was built using a customized polytetrafluoroethylene (PTFE) container and matched with vector network analysis technology for experimental testing. The Gated-Reflect-Line (GRL) calibration technique was adopted to mitigate measurement errors arising from multiple signal reflections and scattering between the container structure and the test samples. Following calibration, liquid or powder samples were loaded into the container, and the reflection S-parameters captured by a vector network analyzer were utilized to calculate the sample dielectric constant. Solid PTFE was tested as a standard sample to verify the feasibility and reliability of the established test system. Further experiments were conducted on quartz powder and dimethyl synthetic silicone oil within the frequency range of 75~110 GHz. The test results show that the maximum deviation of measured data from the average value is merely 5%, which verifies the superior accuracy of the proposed method. This study offers a crucial technical reference for millimeter-wave dielectric constant measurement of liquid and powder materials, supporting technical development in food processing, biomedicine and other related fields.

    参考文献
    相似文献
    引证文献
引用本文

王梦慧, 邹翘, 周治鑫, 董继刚, 王亚海, 赵锐.毫米波频段液体及粉末的介电常数测试技术研究[J].计测技术,2026,46(4):148~155:
10.11823/j. issn.1674-5795.2026.04.11.

复制
分享
相关视频

文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:
  • 最后修改日期:
  • 录用日期:
  • 在线发布日期: 2026-09-09
  • 出版日期:
文章二维码