Determination of the Location for Heat Flux Sensor Inside the Cavity Using In-Cavity Calibration Method
CSTR:
Author:
Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    This paper introduces the methods of heat flux sensor calibration and gives detailed descriptions of in-cavity calibration instruments and principles. The calibration has been performed using a sodium heat pipe blackbody and a high temperature blackbody, both with 50-mm-diam cavities. Heat flux sensors used here are circular foil type and water-cooled. An equivalent cavity model has been established to calculate the effective emissivity. The computations and experimental results show that the optimum location for the sensor is at a distance of about one cavity diameter from the cavity bottom.