Study of Reproducibility of Automatic Freezing and Saving Device in the Water Triple Point Cell
Author:
Affiliation:

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    介绍了型号为 CIMM-TH- 0230的水三相点瓶自动冻制保存装置,对该装置进行水三相点温坪复现性研究,并考核装置的各项指标。实验结果显示: 该装置冻制的水三相点的温坪在48 h 内变化不大于 0.2 mK,复现性小于 0.04 mK,稳定性优于0. 5 mK,不同瓶子之间复现的温度差值不大于0.4 mK。试验数据表明该装置非常适合水三相点瓶的自动冻制? 保存与复现。

    Abstract:

    The paper introduces an automatic freezing and saving device in the water triple point cell for CIMM-TH-0230.The reproducibility of the temperature plateau in the water triple point cell is studied with the device,and the performance of the device is evaluated.The test results show that: the change in the temperature plateau of the triple point of water with the freezing device is not more than 0.2 mK within 48 h;the reproducibility is less than 0.06 mK; the repeatability is better than 0.5 mK;the temperature differences between different cells are not more than 0.4 mK. According to the test results,the device is suitable for freezing,saving and reproducing of the water triple point cell.